Performance evaluation of Automated Fluorescent Immunoassay System ROTA and NORO for detection of rotavirus and norovirus: A comparative study of assay performance with RIDASCREEN® Rotavirus and Norovirus

J Clin Lab Anal. 2021 Jan;35(1):e23585. doi: 10.1002/jcla.23585. Epub 2020 Sep 23.

Abstract

Background: The Automated Fluorescent Immunoassay System ROTA (AFIAS-Rota) and NORO (AFIAS-Noro) assays (Boditech Med Inc.) are newly developed diagnostic tests for rotavirus and norovirus infections.

Methods: Performance of AFIAS-Rota/Noro assays was evaluated in comparison with RIDASCREEN® Rotavirus and Norovirus ELISA kits (R-Biopharm) using clinical stool samples submitted from November 2018 to January 2019. Multiplex real-time reverse transcription-polymerase chain reaction was used as reference method.

Results: A total of 256 clinical specimens were analyzed. AFIAS-Rota and RIDASCREEN Rotavirus had almost perfect agreement (Kappa value = 0.95), and substantial agreement was observed between AFIAS-Noro and RIDASCREEN Norovirus (Kappa value = 0.80). For detection of rotavirus, AFIAS and RIDASCREEN assays showed satisfactory diagnostic sensitivity (100% and 97.8%, respectively) and specificity (99.5% and 99.1%). For detection of norovirus, the RIDASCREEN assay showed significantly higher sensitivity than the AFIAS-Noro (86.0% and 66.0%, respectively; P = .002). Analytic specificity of AFIAS-Rota/Noro assays showed no cross-reactivity against any other bacteria (14 strains) or viruses (2 strains). Hands-on time (6 minutes) and turnaround time (26 minutes) required to perform AFIAS assays were much shorter than those required for RIDASCREEN assays (20 and 150 minutes, respectively).

Conclusion: The AFIAS-Rota/Noro assays showed overall excellent agreement with the RIDASCREEN assays. Although the AFIAS-Noro assay exhibited lower sensitivity than the RIDASCREEN Norovirus assay for detection of norovirus, the AFIAS-Rota/Noro assays could be useful as a rapid initial screening test in clinical laboratories due to its convenience and rapid turnaround time.

Keywords: AFIAS-Noro; AFIAS-Rota; immunoassay; norovirus; performance; rotavirus.

MeSH terms

  • Caliciviridae Infections / diagnosis*
  • Feces / virology
  • Fluorescent Antibody Technique* / methods
  • Fluorescent Antibody Technique* / standards
  • Fluorescent Antibody Technique* / statistics & numerical data
  • Humans
  • Norovirus / genetics
  • Norovirus / immunology
  • Norovirus / isolation & purification*
  • Real-Time Polymerase Chain Reaction
  • Reproducibility of Results
  • Rotavirus / genetics
  • Rotavirus / immunology
  • Rotavirus / isolation & purification*
  • Rotavirus Infections / diagnosis*
  • Sensitivity and Specificity
  • Virology