Nineteen-year outcomes with the Anaconda stent graft system from two tertiary centers

J Vasc Surg. 2021 Jul;74(1):105-113. doi: 10.1016/j.jvs.2020.12.060. Epub 2021 Mar 2.

Abstract

Objective: The Anaconda stent graft device presents peculiar and unique features in its design and deployment sequence, available on the market for many years with minor modifications, among different generations. The aim of this study was to assess the long-term, real-life, outcomes obtained in two high-volume centers with this endograft for infrarenal aortic aneurysms treatment over a period of 19 years.

Methods: All patients treated with the investigated device between 2000 and 2019, with at least 1 year of follow-up, were entered into a prospective database and data were reviewed retrospectively. The long-term results were assessed according to the Kaplan-Meier method. Primary outcomes were overall survival, freedom from AAA rupture, and freedom from AAA-related death. Secondary outcomes were freedom from late reintervention, freedom from late conversion to open repair, freedom from late limb occlusion, and freedom from late EVAR failure.

Results: A total of 260 consecutive patients were treated during the study period. Technical success was achieved in 98.5% (256/260) of the cases. Two perioperative deaths (0.8%) and four major complications (1.5%) were recorded. At a mean follow-up of 54 ± 38 months (summary follow-up index = 0.86), the late survival estimate was 94.8% at 1 year, 75.2% at 5 years, 50.5% at 10 years, and 43.3% at 13 years, with no case of late AAA rupture or late AAA-related death. Late freedom from conversion was 99.2% at 1 year, 96.9% at 5 years, 94.0% at 10 years, and 94.0% at 13 years. Freedom from late limb thrombosis was 96.8% at 1 year and 91.0% at 5, 10, and 13 years. Estimates of freedom from late reintervention was 95.6% at 1 year, 87.0% at 5 years, 77.4% at 10 years, and 77.4% at 13 years. Estimates of limb thrombosis at 11 years were significantly different in patients receiving or not intraoperative iliac kissing balloon maneuver with noncompliant angioplasty catheters (98.0% vs 88.4%; P = .029).

Conclusions: The Anaconda stent graft demonstrated effective and durable outcomes for AAA treatment over a 19-year period, with no case of late AAA rupture or AAA related death in the studied population. Late limb occlusion was the leading cause of reintervention; however, a positive prevention strategy was found in intraoperative kissing balloon, which was associated with significant reduced rate of iliac thrombosis over time.

Keywords: Abdominal aortic aneurysm; Anaconda endograft; Durability; Endovascular; Limb patency; Long-term results.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Aneurysm, Abdominal / diagnostic imaging
  • Aortic Aneurysm, Abdominal / mortality
  • Aortic Aneurysm, Abdominal / physiopathology
  • Aortic Aneurysm, Abdominal / surgery*
  • Blood Vessel Prosthesis Implantation / adverse effects
  • Blood Vessel Prosthesis Implantation / instrumentation*
  • Blood Vessel Prosthesis Implantation / mortality
  • Blood Vessel Prosthesis*
  • Databases, Factual
  • Endovascular Procedures / adverse effects
  • Endovascular Procedures / instrumentation*
  • Endovascular Procedures / mortality
  • Female
  • Graft Occlusion, Vascular / etiology
  • Graft Occlusion, Vascular / physiopathology
  • Hospitals, High-Volume
  • Humans
  • Italy
  • Male
  • Middle Aged
  • Progression-Free Survival
  • Prosthesis Design
  • Retrospective Studies
  • Risk Assessment
  • Risk Factors
  • Stents*
  • Tertiary Care Centers
  • Time Factors
  • Vascular Patency