A multi-channel x-ray temporal diagnostic for measurement of time-resolved electron temperature in cryogenic deuterium-tritium implosions at OMEGA

Rev Sci Instrum. 2021 Feb 1;92(2):023507. doi: 10.1063/5.0042329.

Abstract

Electron-temperature (Te) measurements in implosions provide valuable diagnostic information, as Te is unaffected by residual flows and other non-thermal effects unlike ion temperature inferred from a fusion product spectrum. In OMEGA cryogenic implosions, measurement of Te(t) can be used to investigate effects related to time-resolved hot-spot energy balance. The proposed diagnostic utilizes five fast-rise (∼15 ps) scintillator channels with distinct x-ray filtering. Titanium and stepped aluminum filtering were chosen to maximize detector sensitivity in the 10 keV-20 keV range, as it has been shown that these x rays have similar density and temperature weighting to the emitted deuterium-tritium fusion neutrons. Initial data collected using a prototype nosecone on the existing neutron temporal diagnostic demonstrate the validity of this diagnostic technique. The proposed system will be capable of measuring spatially integrated Te(t) with 20 ps time resolution and <10% uncertainty at peak emission in cryogenic DT implosions.