Objectives: The accuracy of point-of-care blood glucometers in the detection and evaluation of neonatal hypoglycemia is a concern. This study compared the performance of three i-SENS glucometers with that of the YSI 2300 STAT Plus Analyzer, which was used as a reference.
Methods: The leftover neonatal capillary blood samples of 319 patients were used in this study. The evaluation process and accuracy performance criteria were based on the International Organization for Standardization 15197:2013 guidelines. The evaluation involved three i-SENS glucometers (BAROzen H Expert plus, CareSens PRO, and CareSens H Beat) and the ACCU-CHEK® Inform II glucometer.
Results: The accuracy evaluation yielded acceptable results as follows: a) 100 and 100% for BAROzen H Expert plus; 99.8 and 100% for CareSens PRO; 98.7%, and 97.2% for CareSens H Beat glucometers were within the range of ±0.8 mmol/L (15 mg/dL) and ±15% of the average reference values at glucose concentrations <5.55 mmol/L (100 mg/dL) and ≥5.55 mmol/L (100 mg/dL), respectively, and b) all estimated glucose values (100%) were within the zones A and B of Consensus Error Grid for all three i-SENS glucometers. There was good correlation between the glucose values estimated by the glucometers and the reference values (R>0.990).
Conclusions: This study demonstrated that i-SENS glucometers exhibit acceptable performance and can be used as effective point-of-care devices in neonates.
Keywords: capillary blood glucose levels; hypoglycemia; i-SENS glucometers; neonatal care; neonates; point-of care blood glucometers.