Performance Characteristics of Long Axial Field-of-View PET Scanners with Axial Gaps

IEEE Trans Radiat Plasma Med Sci. 2021 May;5(3):322-330. doi: 10.1109/trpms.2020.3027257. Epub 2020 Sep 28.

Abstract

The introduction of long (>60 cm) axial field-of-view (LAFOV) PET systems has shown their potential for clinical and research applications. LAFOV scanners are expensive, so there is interest in designing systems with longer axial coverage while mitigating cost by introducing detector gaps. We used measurements on the PennPET Explorer (64-cm AFOV prototype) and simulations of scanners up to 143-cm long to assess scanner performance with axial gaps introduced by varying the number of detector rows in each ring. Removing detectors reduces the total sensitivity and results in a non-uniform axial noise profile. Axial resolution shows small (<0.5 mm) loss from the edge of the AFOV to the center, even for a 143-cm AFOV scanner with an unrestricted acceptance angle. The presence of large axial gaps increases the variability in axial resolution and contrast recovery across the AFOV compared to a system without gaps. More modest axial gaps show less variable behavior. The results suggest that designs where the gap is no larger than one-half of the width of a detector ring may be preferred, although the optimal choice of scanner design with the trade-offs of performance and AFOV will depend on its intended usage.

Keywords: Long axial field of view PET; PET scanner design; PennPET Explorer.