Real evidence to assess clinical testing interference risk (REACTIR): A strategy using real world data to assess the prevalence of interfering substances in patients undergoing clinical laboratory testing

Clin Chim Acta. 2021 Dec:523:178-184. doi: 10.1016/j.cca.2021.09.001. Epub 2021 Sep 6.

Abstract

Introduction: Laboratory test interferences can cause spurious test results and patient harm. Knowing the frequency of various interfering substances in patient populations likely to be tested with a particular laboratory assay may inform test development, test utilization and strategies to mitigate interference risk.

Methods: We developed REACTIR (Real Evidence to Assess Clinical Testing Interference Risk), an approach using real world data to assess the prevalence of various interfering substances in patients tested with a particular type of assay. REACTIR uses administrative real world data to identify and subgroup patient cohorts tested with a particular laboratory test and evaluate interference risk.

Results: We demonstrate the application REACTIR to point of care (POC) blood glucose testing. We found that exposure to several substances with the potential to interfere in POC blood glucose tests, including N-acetyl cysteine (NAC) and high dose vitamin C was uncommon in most patients undergoing POC glucose tests with several key exceptions, such as burn patients receiving high dose IV-vitamin C or acetaminophen overdose patients receiving NAC.

Conclusions: Findings from REACTIR may support risk mitigation strategies including targeted clinician education and clinical decision support. Likewise, adaptations of REACTIR to premarket assay development may inform optimal assay design and assessment.

Keywords: Data Science; Glucose; Interference; Laboratory Analytics; Point of care; Vitamin C.

MeSH terms

  • Blood Glucose*
  • Humans
  • Laboratories, Clinical
  • Point-of-Care Systems*
  • Point-of-Care Testing
  • Prevalence

Substances

  • Blood Glucose