Evaluating a Convolutional Neural Network Noise Reduction Method When Applied to CT Images Reconstructed Differently Than Training Data

J Comput Assist Tomogr. 2021 Jul-Aug;45(4):544-551. doi: 10.1097/RCT.0000000000001150.

Abstract

Objective: The aim of this study was to evaluate a narrowly trained convolutional neural network (CNN) denoising algorithm when applied to images reconstructed differently than training data set.

Methods: A residual CNN was trained using 10 noise inserted examinations. Training images were reconstructed with 275 mm of field of view (FOV), medium smooth kernel (D30), and 3 mm of thickness. Six examinations were reserved for testing; these were reconstructed with 100 to 450 mm of FOV, smooth to sharp kernels, and 1 to 5 mm of thickness.

Results: When test and training reconstruction settings were not matched, there was either reduced denoising efficiency or resolution degradation. Denoising efficiency was reduced when FOV was decreased or a smoother kernel was used. Resolution loss occurred when the network was applied to an increased FOV, sharper kernel, or decreased image thickness.

Conclusions: The CNN denoising performance was degraded with variations in FOV, kernel, or decreased thickness. Denoising performance was not affected by increased thickness.

MeSH terms

  • Algorithms
  • Deep Learning
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Neural Networks, Computer*
  • Signal-To-Noise Ratio*
  • Tomography, X-Ray Computed / methods*