Purpose: Well-designed routine multileaf collimator (MLC) quality assurance (QA) is important to assure external-beam radiation treatment delivery accuracy. This study evaluates the clinical necessity of a comprehensive weekly (C-Weekly) MLC QA program compared to the American Association of Physics in Medicinerecommended weekly picket fence test (PF-Weekly), based on our seven-year experience with weekly MLC QA.
Methods: The C-Weekly MLC QA program used in this study includes 5 tests to analyze: (1) absolute MLC leaf position; (2) interdigitation MLC leaf position; (3) picket fence MLC leaf positions at static gantry angle; (4) minimum leaf-gap setting; and (5) volumetric-modulated arc therapy delivery. A total of 20,226 QA images from 16,855 tests (3,371 tests × 5) for 11 linacs at 5 photon clinical sites from May 2014 to June 2021 were analyzed. Failure mode and effects analysis was performed with 5 failure modes related to the 5 tests. For each failure mode, a risk probability number (RPN) was calculated for a C-Weekly and a PF-Weekly MLC QA program. The probability of occurrence was evaluated from statistical analyses of the C-Weekly MLC QA.
Results: The total number of failures for these 16,855 tests was 143 (0.9%): 39 (27.3%) for absolute MLC leaf position, 13 (9.1%) for interdigitation position, 9 (6.3%) for static gantry picket fence, 2 (1.4%) for minimum leaf-gap setting, and 80 (55.9%) for VMAT delivery. RPN scores for PF-Weekly MLC QA ranged from 60 to 192 and from 48 to 96 for C-Weekly MLC QA.
Conclusion: RPNs for the 5 failure modes of MLC QA tests were quantitatively determined and analyzed. A comprehensive weekly MLC QA is imperative to lower the RPNs of the 5 failure modes to the desired level (<125); those from the PF-Weekly MLC QA program were found to be higher (>125). This supports the clinical necessity for comprehensive weekly MLC QA.
Keywords: American Association of Physics in Medicine Task Group 142; electronic portal imaging device; failure mode and effects analysis; leaf-end leakage; multileaf collimator quality assurance; picket fence; volumetric-modulated arc therapy.
© 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine.