Agreement of a Novel Artificial Intelligence Software With Optical Coherence Tomography and Manual Grading of the Optic Disc in Glaucoma

J Glaucoma. 2023 Apr 1;32(4):280-286. doi: 10.1097/IJG.0000000000002147. Epub 2022 Nov 28.

Abstract

Prcis: The offline artificial intelligence (AI) on a smartphone-based fundus camera shows good agreement and correlation with the vertical cup-to-disc ratio (vCDR) from the spectral-domain optical coherence tomography (SD-OCT) and manual grading by experts.

Purpose: The purpose of this study is to assess the agreement of vCDR measured by a new AI software from optic disc images obtained using a validated smartphone-based imaging device, with SD-OCT vCDR measurements, and manual grading by experts on a stereoscopic fundus camera.

Methods: In a prospective, cross-sectional study, participants above 18 years (Glaucoma and normal) underwent a dilated fundus evaluation, followed by optic disc imaging including a 42-degree monoscopic disc-centered image (Remidio NM-FOP-10), a 30-degree stereoscopic disc-centered image (Kowa nonmyd WX-3D desktop fundus camera), and disc analysis (Cirrus SD-OCT). Remidio FOP images were analyzed for vCDR using the new AI software, and Kowa stereoscopic images were manually graded by 3 fellowship-trained glaucoma specialists.

Results: We included 473 eyes of 244 participants. The vCDR values from the new AI software showed strong agreement with SD-OCT measurements [95% limits of agreement (LoA)=-0.13 to 0.16]. The agreement with SD-OCT was marginally better in eyes with higher vCDR (95% LoA=-0.15 to 0.12 for vCDR>0.8). Interclass correlation coefficient was 0.90 (95% CI, 0.88-0.91). The vCDR values from AI software showed a good correlation with the manual segmentation by experts (interclass correlation coefficient=0.89, 95% CI, 0.87-0.91) on stereoscopic images (95% LoA=-0.18 to 0.11) with agreement better for eyes with vCDR>0.8 (LoA=-0.12 to 0.08).

Conclusions: The new AI software vCDR measurements had an excellent agreement and correlation with the SD-OCT and manual grading. The ability of the Medios AI to work offline, without requiring cloud-based inferencing, is an added advantage.

MeSH terms

  • Artificial Intelligence
  • Cross-Sectional Studies
  • Glaucoma* / diagnosis
  • Humans
  • Intraocular Pressure
  • Optic Disk*
  • Optic Nerve Diseases* / diagnosis
  • Photography / methods
  • Prospective Studies
  • Reproducibility of Results
  • Software
  • Tomography, Optical Coherence / methods