Background: Most randomized studies testing the effectiveness of IVBT were limited to vessels less than 4 mm diameter. In fact, it is now common to treat vessels larger than 4 mm. Accordingly, the authors instituted a prescription dose increase to 34 Gy at 2 mm from source center for vessels greater than 4.0 mm. The increase in prescription dose to 34 Gy at 2 mm from center is substantial, being 50% higher than the conventional maximum of 23 Gy.
Aim: To take a close look at group of patients treated to 34 Gy, and for whom follow-up angiograms are available.
Methods: Ten patients treated for ISR with a prescription dose of 34 Gy and for whom follow-up angiograms were available were studied. Beta-radiation brachytherapy was performed with a Novoste Beta-Cath System using a strontium-90 (beta) source (Best Vascular, Springfield, VA). Source lengths of 40 or 60 mm were used. A dose of 34 Gy was prescribed at 2 mm from the source center.
Results: Patients were re-catheterized from 2 to 21 months (median: 16 months) following IVBT, all for symptoms suggested of restenosis. All patients had some degree of ISR of the target vessel, but no IVBT-treated vascular segment showed angiographic signs of degeneration, dissection or aneurysm.
Conclusion: The authors' clinical impression, along with detailed review of the 10 cases, suggest that using a 34 Gy prescription dose at 2 mm from source center does not result in increased toxicity.
Keywords: brachytherapy; cardiac; coronary; dosimetry; radiation; restenosis.
© 2023 Wiley Periodicals LLC.