Effect of storage time of paraffin sections on the expression of PD-L1 (SP142) in invasive breast cancer

Diagn Pathol. 2023 Dec 5;18(1):131. doi: 10.1186/s13000-023-01423-8.

Abstract

Background: PD-L1 staining using long-stored paraffin sections may not be consistent with the true PD-L1 expression of patients. Therefore, it is necessary to explore the expression of PD-L1(SP142) in paraffin sections of invasive breast cancer with different storage times and the optimal storage temperature for unstained paraffin sections.

Methods: The study included 71 cases of PD-L1(SP142) positive breast cancer. The unstained paraffin sections were stored at room temperature conditions (20-25 °C), 4 °C, -20 °C and - 80 °C, respectively. PD-L1 staining was performed at 1, 2, 3, 4, 8, 12 and 24 weeks of storage. PD-L1 expression was assessed with a continuity score.

Results: The PD-L1 antigen was gradually lost as the storage time of paraffin sections increased. The PD-L1 positivity rate was 97.18% at 1 week for the sections stored at room temperature, and decreased from 83.10 to 71.83% for the sections stored for 2 weeks to 4 weeks, and 61.97%, 54.93%, and 32.93% for 8, 12, and 24 weeks, respectively. When stored at low temperatures of 4 °C, -20 °C and - 80 °C, the positivity rate decreases with the same trend but more slowly compared to room temperature. The mean IC score of PD-L1 also showed a gradual decrease in all cases. In the consistency analysis, PD-L1 expression in slices stored at room temperature for 2 weeks was consistent with PD-L1 expression in fresh slices (ICC ≥ 0.9 for all slices), and PD-L1 expression in slices stored at 4 °C or -20 °C for 4 weeks was consistent with PD-L1 expression in fresh slices (ICC ≥ 0.9 for all slices). When stored under refrigeration at -80 °C, PD-L1 expression in slices stored for 3 weeks was consistent with that in fresh slices (ICC ≥ 0.9).

Conclusions: To our knowledge, this is the first article on the effect of preservation time and preservation temperature of paraffin sections on PD-L1 expression in breast cancer. Long-term storage of paraffin sections of unstained invasive breast cancer can lead to antigen loss of PD-L1 (SP142). Refrigerated storage of paraffin sections can delay antigen loss, with best results at 4 °C or -20 °C, and a storage time of no more than 4 weeks is recommended.

Keywords: Breast cancer; Immunohistochemistry; PD-L1; Paraffin sections; Storage time.

MeSH terms

  • B7-H1 Antigen / metabolism
  • Biomarkers, Tumor / analysis
  • Breast Neoplasms*
  • Female
  • Humans
  • Immunohistochemistry
  • Paraffin
  • Time Factors