Asinina de Miranda is a protected donkey sub-species from the Mirandês plateau in northeastern of Portugal. Donkeys are animals that have substantially lost their place as working animals in modern society, this had led to a decrease in their population numbers. A need to preserve native species has led to the foundation of organizations like Associação para o Estudo e Proteção do Gado Asinino (AEPGA) and the development of studies regarding breed welfare, such as hematology. The IDEXX ProCyte Dx is a veterinary hematology analyzer validated for several species, but not for donkeys. The aim of this study was to validate the ProCyte Dx for Asinina de Miranda donkeys. The validation requires a controlled study of precision, carryover, linearity and comparison between the equipment and the manually obtained values for the leukocyte differential count and hematocrit. Results indicated coefficient of variation was good (below 5 %) for both the intra-assay and the inter-assay precision, except for basophils. Carryover was 0 % for all the parameters except platelets (5.88 %). Linearity showed a very high Pearson correlation coefficient, above 0.99, for erythrocytes, hematocrit, hemoglobin, leucocytes, neutrophils, lymphocytes, monocytes, eosinophils, platelets and plateletcrit. Comparison demonstrated excellent agreement for hematocrit (rs=0.96) and good Spearman rank correlation for neutrophils (rs=0.84) and lymphocytes (rs=0.90). Accuracy for total leukocyte count and platelets could not be determined. In conclusion, the ProCyte Dx seems appropriate to be used in Asinina de Miranda hematology.
Keywords: Asinina de Miranda; Donkey; Hematology; ProCyte Dx (IDEXX); Validation.
Copyright © 2023 Elsevier Inc. All rights reserved.