Intra-Abdominal Hemorrhage Triggering Inappropriate Therapy From a Subcutaneous Defibrillator

JACC Case Rep. 2023 Dec 20;29(3):102167. doi: 10.1016/j.jaccas.2023.102167. eCollection 2024 Feb 7.

Abstract

The SMART Pass filter (Boston Scientific) aims to reduce inappropriate shocks (IASs) from subcutaneous implantable cardioverter-defibrillators by filtering out low-frequency signals such as T waves. However, this filter is deactivated in the presence of diminished R-wave sensing. We describe a case of IAS in the setting of extensive intra-abdominal hemorrhage.

Keywords: SMART Pass algorithm; inappropriate shock; subcutaneous ICD.