Feasibility and clinical implementation of MRI-guided surface brachytherapy

J Contemp Brachytherapy. 2024 Feb;16(1):48-56. doi: 10.5114/jcb.2024.135647. Epub 2024 Feb 23.

Abstract

Purpose: Best practices for high-dose-rate surface applicator brachytherapy treatment (SABT) have long relied on computed tomography (CT)-based imaging to visualize diseased sites for treatment planning. Compared with magnetic resonance (MR)-based imaging, CT provides insufficient soft tissue contrast. This work described the feasibility of clinical implementation of MR-based imaging in SABT planning to provide individualized treatment optimization.

Material and methods: A 3D-printed phantom was used to fit Freiberg flap-style (Elekta, The Netherlands) applicator. Images were taken using an optimized pointwise encoding time reduction with radial acquisition (PETRA) MR sequence for catheter visualization, and a helical CT scan to generate parallel treatment plans. This clinical study included three patients undergoing SABT for Dupuytren's contracture/palmar fascial fibromatosis imaged with the same modalities.SABT planning was performed in Oncentra Brachy (Elekta Brachytherapy, The Netherlands) treatment planning software. A geometric analysis was conducted by comparing CT-based digitization with MR-based digitization. CT and MR dwell positions underwent a rigid registration, and average Euclidean distances between dwell positions were calculated. A dosimetric comparison was performed, including point-based dose difference calculations and volumetric segmentations with Dice similarity coefficient (DSC) calculations.

Results: Euclidean distances between dwell positions from CT-based and MR-based plans were on average 0.68 ±0.05 mm and 1.35 ±0.17 mm for the phantom and patients, respectively. The point dose difference calculations were on average 0.92% for the phantom and 1.98% for the patients. The D95 and D90 DSC calculations were both 97.9% for the phantom, and on average 93.6% and 94.2%, respectively, for the patients.

Conclusions: The sub-millimeter accuracy of dwell positions and high DSC's (> 0.95) of the phantom demonstrated that digitization was clinically acceptable, and accurate treatment plans were produced using MR-only imaging. This novel approach, MRI-guided SABT, will lead to individualized prescriptions for potentially improved patient outcomes.

Keywords: brachytherapy; high-dose-rate; magnetic resonance; surface brachytherapy.