Aim The efficacy of hybrid closed-loop systems (HCLs) in managing glycemic control in pregnant women with type 1 diabetes remains inadequately characterized. We evaluated the use of the Medtronic Minimed 780G HCLs.
Methods: The retrospective observational study analyzed the glycemic and perinatal outcomes of pregnant women using the HCLs, followed at our tertiary centre. Independent t-tests were employed to compare data among trimesters based on pre-pregnancy HbA1c. The associations between glycemic parameters and perinatal outcomes were explored using Spearman rho.
Results: Among the 21 women (age: 33.5 ± 4.2 years, diabetes duration: 21.2 ± 7.6 years, pre-pregnancy HbA1c 7.0 ± 1.1 % (52.9 ± 11.9 mmol/mol)) time in range (pTIR, 63-140 mg/dl; 3.5-7.8 mmol/l) increased progressively throughout pregnancy (trimesters: first: 64.0 ± 9.0 %; second:71.3 ± 11.8 %; third: 75.7 ± 8.1 %). Simultaneously, mean sensor glucose decreased (trimesters: first: 130 ± 10.4 mg/dl (7.2 ± 0.6 mmol/l); second: 120.9 ± 13.4 mg/dl (6.7 ± 0.7 mmol/l); third: 117.3 ± 9.1 mg/dl (6.5 ± 0.5 mmol/l)). Although a majority of women achieved the target pTIR until the third trimester, this did not consistently prevent the delivery of a large-for-gestational-age baby. Notably, one ketoacidosis event occurred, and there were no reported instances of severe hypoglycemia.
Conclusion: Use of the Minimed 780G HCLs enabled the attainment of recommended pregnancy glycemic targets for most women with type 1 diabetes in a real-world setting.
Keywords: Automated insulin delivery; Closed-loop insulin delivery; Continuous glucose monitoring; Pregnancy; Pregnancy outcomes; Time in range; Type 1 diabetes.