Reintervention Rate after Treatment with the INCRAFT AAA Ultra-Low-Profile Stent Graft System

Ann Vasc Surg. 2024 Dec:109:216-224. doi: 10.1016/j.avsg.2024.05.023. Epub 2024 Jul 15.

Abstract

Background: The INCRAFT stent graft system is an ultra-low-profile endograft for the exclusion of infrarenal aortic aneurysms. In the market approval studies, an increased rate of device-related complications was observed and the endograft was approved with mandated postmarketing investigations. Our aim was to analyze mid-term outcomes of a real-world patient cohort treated with the INCRAFT endograft.

Methods: Consecutive patients treated with the INCRAFT endograft between February 2015 and December 2022 at a single institution were included. In accordance with the Society for Vascular Surgery reporting standards, safety endpoints were reported and outcome endpoints included reinterventions, technical success, aortic-related and overall-mortality, endoleak, stent fracture, and endograft migration >5 mm.

Results: Eighty patients (85% male) with a mean age of 76 ± 7 years were included. Fifty-two patients (65%) were treated within the endograft's instruction for use. Mean aortic diameter was 59 ± 10 mm and 91% of the procedures were performed percutaneously. Mean follow-up (FU) was 37 ± 25 months and there was no aortic- or procedure-related mortality. Reinterventions occurred in 25 patients (31%) with a freedom from reintervention at 1, 3, and 5 years of 84%, 66%, and 55%. The most frequent reinterventions were limb graft stenting (23%) and type II endoleak embolization (14%). Limb occlusion rate was 9% and in 3 patients (4%) distal endograft migrations >5 mm occurred. Persisting type II endoleaks were observed in 29% and aneurysm diameter was stable in 41% and had shrunk in 38%. Three type III endoleaks (4%) developed during FU and 4 open conversions (5%) were necessary. No known risk factors, including treatment outside instruction for use, were predictive for reinterventions.

Conclusions: Treatment of infrarenal aortic aneurysms with the INCRAFT stent graft system was safe and successful. Nevertheless, a substantial rate of reinterventions was necessary during FU to maintain endograft patency and prevent aneurysm growth.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Aneurysm, Abdominal* / diagnostic imaging
  • Aortic Aneurysm, Abdominal* / mortality
  • Aortic Aneurysm, Abdominal* / surgery
  • Blood Vessel Prosthesis Implantation* / adverse effects
  • Blood Vessel Prosthesis Implantation* / instrumentation
  • Blood Vessel Prosthesis Implantation* / mortality
  • Blood Vessel Prosthesis*
  • Endoleak / etiology
  • Endovascular Procedures* / adverse effects
  • Endovascular Procedures* / instrumentation
  • Endovascular Procedures* / mortality
  • Female
  • Foreign-Body Migration / etiology
  • Humans
  • Male
  • Postoperative Complications / etiology
  • Postoperative Complications / mortality
  • Prosthesis Design*
  • Prosthesis Failure
  • Retreatment
  • Retrospective Studies
  • Risk Assessment
  • Risk Factors
  • Stents*
  • Time Factors
  • Treatment Outcome