Optimal image quality and radiation doses with optimal tube voltages/currents for pediatric anthropomorphic phantom brains

PLoS One. 2024 Jul 22;19(7):e0306857. doi: 10.1371/journal.pone.0306857. eCollection 2024.

Abstract

Objective: Using pediatric anthropomorphic phantoms (APs), we aimed to determine the scanning tube voltage/current combinations that could achieve optimal image quality and avoid excessive radiation exposure in pediatric patients.

Materials and methods: A 64-slice scanner was used to scan a standard test phantom to determine the volume CT dose indices (CTDIvol), and three pediatric anthropomorphic phantoms (APs) with highly accurate anatomy and tissue-equivalent materials were studied. These specialized APs represented the average 1-year-old, 5-year-old, and 10-year-old children, respectively. The physical phantoms were constructed with brain tissue-equivalent materials having a density of ρ = 1.07 g/cm3, comprising 22 numbered 2.54-cm-thick sections for the 1-year-old, 26 sections for the 5-year-old, and 32 sections for the 10-year-old. They were scanned to acquire brain CT images and determine the standard deviations (SDs), effective doses (EDs), and contrast-to noise ratios (CNRs). The APs were scanned by 21 combinations of tube voltages/currents (80, 100, or 120 kVp/10, 40, 80, 120, 150, 200, or 250 mA) and rotation time/pitch settings of 1 s/0.984:1.

Results: The optimal tube voltage/current combinations yielding optimal image quality were 80 kVp/80 mA for the 1-year-old AP; 80 kVp/120 mA for the 5-year-old AP; and 80 kVp/150 mA for the 10-year-old AP. Because these scanning tube voltages/currents yielded SDs, respectively, of 12.81, 13.09, and 12.26 HU, along with small EDs of 0.31, 0.34, and 0.31 mSv, these parameters and the induced values were expediently defined as optimal.

Conclusions: The optimal tube voltages/currents that yielded optimal brain image quality, SDs, CNRs, and EDs herein are novel and essentially important. Clinical translation of these optimal values may allow CT diagnosis with low radiation doses to children's heads.