Multicenter Pivotal Study of the Alterra Adaptive Prestent for the Treatment of Pulmonary Regurgitation

JACC Cardiovasc Interv. 2024 Oct 14;17(19):2287-2297. doi: 10.1016/j.jcin.2024.07.036.

Abstract

Background: A dilated native right ventricular outflow tract (RVOT) presents unique challenges for transcatheter management using balloon-expandable valves. The Alterra Adaptive Prestent was designed to expand transcatheter therapy to treat patients with dilated RVOTs.

Objectives: The aim of this study was to report 2-year outcomes of the main cohort of the ALTERRA (Multicenter Study of Congenital Pulmonic Valve Dysfunction Studying the SAPIEN 3 THV With the Alterra Adaptive Prestent) pivotal trial using the prestent with transcatheter pulmonary valve replacement.

Methods: The prestent device used with the 29 mm SAPIEN 3 transcatheter heart valve (THV) was evaluated for the management of patients with moderate or greater pulmonary valve regurgitation (PR). The primary endpoint was THV dysfunction at 6 months, defined as a nonhierarchical composite of RVOT/pulmonary valve reintervention, moderate or greater total PR on transthoracic echocardiography, and mean RVOT/pulmonary valve gradient 35 mm Hg or greater on transthoracic echocardiography. The primary endpoint and outcomes through 2 years are presented in this analysis.

Results: Of 97 patients screened, 60 underwent prestent and THV implantation. There was 1 staged procedure. No patients had THV dysfunction at 6 months. At 2 years, the majority of patients (92.5%) had mild or less PR, with no reports of coronary compression, stent fractures warranting reintervention, or endocarditis. Of the 21 patients (34.4%) who experienced early (days 0-1) arrhythmias, 12 had episodes of nonsustained ventricular tachycardia that resolved with medication. One patient underwent reintervention secondary to an iatrogenic RVOT obstruction; there were no deaths or explantations through 2 years.

Conclusions: The Alterra prestent in combination with the SAPIEN 3 THV has excellent outcomes at 2 years, with no significant valve dysfunction in the main pivotal cohort.

Keywords: Alterra Adaptive Prestent; SAPIEN 3 THV; pulmonary regurgitation; right ventricular outflow tract/pulmonary valve dysfunction; transcatheter pulmonary valve replacement.

Publication types

  • Multicenter Study
  • Clinical Trial

MeSH terms

  • Adolescent
  • Adult
  • Balloon Valvuloplasty / adverse effects
  • Cardiac Catheterization* / adverse effects
  • Cardiac Catheterization* / instrumentation
  • Europe
  • Female
  • Heart Valve Prosthesis Implantation* / adverse effects
  • Heart Valve Prosthesis Implantation* / instrumentation
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Prosthesis Design*
  • Pulmonary Valve Insufficiency* / diagnostic imaging
  • Pulmonary Valve Insufficiency* / etiology
  • Pulmonary Valve Insufficiency* / physiopathology
  • Pulmonary Valve Insufficiency* / surgery
  • Pulmonary Valve* / diagnostic imaging
  • Pulmonary Valve* / physiopathology
  • Pulmonary Valve* / surgery
  • Recovery of Function*
  • Risk Factors
  • Time Factors
  • Treatment Outcome
  • United States
  • Young Adult