Objectives: A microfluidic flow cytometer-based point-of-care (POC) analyzer was validated against an in-laboratory hematology analyzer (Sysmex XN Automated Hematology System). Concordance on a full complete blood cell count (CBC) with 5-part differential, as performed by operators with no prior clinical laboratory experience, was evaluated.
Methods: We prospectively collected 376 venous blood specimens (376) from individuals with self-reported medical conditions and from apparently healthy individuals. Forty-six additional remnant specimens were acquired to ensure coverage of analytic measuring ranges. Parallel testing was performed, with up to 7 hours between testing on the POC and Sysmex XN analyzers.
Results: Regression analysis resulted in r values of 0.998 to 0.932 for all parameters of a 5-part differential CBC other than basophils (0.709). The mean percentage bias from the reference method, inclusive of the upper and lower reporting limits, was less than 2% for parameters other than lymphocytes (-6.4%), monocytes (25.9%), eosinophils (12.2%), and basophils (-15%). Overall agreement on abnormal flagging was 93.3%.
Conclusions: The Cito CBC microflow cytometer (CytoChip Inc) provides a CBC with a 5-part differential with accuracy, precision, and abnormal flagging equivalent to a moderate-complexity hematology analyzer. It has the key features required of a POC device that can be operated in a waived setting: minimum space requirements, rapid results, single-action measurement (no sample processing or dilution), ease of use, and minimal blood volume.
Keywords: 5-part differential; CBC; flow cytometer; microfluidic; point of care.
© The Author(s) 2024. Published by Oxford University Press on behalf of American Society for Clinical Pathology. All rights reserved. For commercial re-use, please contact reprints@oup.com for reprints and translation rights for reprints. All other permissions can be obtained through our RightsLink service via the Permissions link on the article page on our site—for further information please contact journals.permissions@oup.com.