Background: Although routine monitoring is not needed for DOACs, knowing if a clinically relevant DOAC level is present can be critical, especially in cases of severe bleeding or urgent surgery. Rapid assays to exclude these levels are necessary but not widely available.
Methods: The MRX PT DOAC assay measures the functional effect of DOACs using the clot-time ratio; a ratio between DOAC-sensitive prothrombin time (PT) and DOAC-insensitive PT. We conducted a multi-center retrospective study of 152 samples from 151 patients with known DOAC levels to assess whether the MRX PT DOAC assay could exclude clinically relevant drug levels of >50 ng/mL, and whether test performance differed across coagulation analyzers. To assess generalizability, the assay was run on four coagulation analyzers: Werfen ACLTOP 750, Diagnostica Stago STACompact MAX, Sysmex CS2500, and Sysmex CN-6000.
Results: The MRX PT DOAC assay had a sensitivity of 100% with a confidence interval (CI) of 70-100% and negative predictive value (NPV) of 100% (CI:57-100%) for edoxaban drug levels >50 ng/mL. For rivaroxaban, sensitivity was 100% (CI:61-100%) and NPV was 100% (CI:5-100%). For apixaban, sensitivity ranged from 59% to 83% (CI:41-93%) and NPV ranged from 0% to 50% (CI:0-69%). The specificity of the assay ranged from 61% to 86% (CI:39-93%) for apixaban, 36% to 50% (CI:3-95%) for edoxaban, and 75% to 100% (CI:30-100%) for rivaroxaban.
Conclusion: The MRX PT DOAC assay reliably excludes clinically relevant DOAC levels edoxaban and rivaroxaban, across multiple analyzers, but not for apixaban.
Keywords: DOAC; anticoagulation; reversal.
Crown Copyright © 2024. Published by Elsevier Inc. All rights reserved.