Accurate diagnosis of ischemic heart disease without exposure to radiation using non-stress unshielded magnetocardiography

Am Heart J Plus. 2024 Nov 14:49:100483. doi: 10.1016/j.ahjo.2024.100483. eCollection 2025 Jan.

Abstract

Study objectives: To evaluate the capability and accuracy of magnetocardiography (MCG) to identify patients with ischemic chest pain from those with non-ischemic pain and to verify normalcy in the MCG in healthy subjects.

Design: We studied 133 patients (mean age 59 ± 14 years, 69 % male) with chronic or acute chest pain syndrome and 63 healthy subjects (mean age 41.7 ± 12.2 years, 51 % male) using unshielded cryogenically cooled MCG systems (Cardiomag Imaging Inc., 9 and 36 channels) in a general clinical setting. Scan time was 90 s to 6 min. Interventions: The MCG data were processed with the same automated analysis software and results were immediately available. All patients were chest pain free at the time of scanning.

Results: A diagnosis of ischemic chest pain was established in 41 % after non-invasive and invasive testing. Rest MCG was normal in all healthy subjects. An abnormal rest MCG was strongly associated with ischemic chest pain, p < 0.0001 (sensitivity of 86 %, specificity of 80 %, positive (PPV) and negative predictive value (NPV) of 75 % and 89 %, respectively). In comparison, the sensitivity, specificity, PPV and NPV of stress SPECT was 93 %, 72 %, 77 % and 91 %, respectively.

Conclusion: Resting MCG is a rapid risk-free method for the detection of ischemic chest pain without the use of radiation or contrast with results comparable with stress SPECT.

Keywords: Angina; Coronary artery disease; Ischemia; Magnetocardiography.