Fully automated methods for the determination of hydrochlorothiazide in human plasma and urine

J Pharm Biomed Anal. 1994 Dec;12(12):1555-62. doi: 10.1016/0731-7085(94)00096-4.

Abstract

LC assays utilizing fully automated sample preparation procedures on Zymark PyTechnology Robot and BenchMate Workstation for the quantification of hydrochlorothiazide (HCTZ) in human plasma and urine have been developed. After aliquoting plasma and urine samples, and adding internal standard (IS) manually, the robot executed buffer and organic solvent addition, liquid-liquid extraction, solvent evaporation and on-line LC injection steps for plasma samples, whereas, BenchMate performed buffer and organic solvent addition, liquid-liquid and solid-phase extractions, and on-line LC injection steps for urine samples. Chromatographic separations were carried out on Beckman Octyl Ultrasphere column using the mobile phase composed of 12% (v/v) acetonitrile and 88% of either an ion-pairing reagent (plasma) or 0.1% trifluoroacetic acid (urine). The eluent from the column was monitored with UV detector (271 nm). Peak heights for HCTZ and IS were automatically processed using a PE-Nelson ACCESS*CHROM laboratory automation system. The assays have been validated in the concentration range of 2-100 ng ml-1 in plasma and 0.1-20 micrograms ml-1 in urine. Both plasma and urine assays have the sensitivity and specificity necessary to determine plasma and urine concentrations of HCTZ from low dose (6.25/12.5 mg) administration of HCTZ to human subjects in the presence or absence of losartan.

MeSH terms

  • Chromatography, Liquid / methods*
  • Humans
  • Hydrochlorothiazide / blood*
  • Hydrochlorothiazide / urine*
  • Research Design
  • Robotics
  • Spectrophotometry, Ultraviolet

Substances

  • Hydrochlorothiazide