MRI scanner transport system

Magn Reson Med. 1994 Jul;32(1):133-6. doi: 10.1002/mrm.1910320119.

Abstract

A computer-controlled transport system has been designed and tested for small bore magnetic resonance scanners for animal research. The transport system allows remote animal positioning based on scout images using a graphic interface. The design utilized commercially available components suitable for use in the scanner's strong magnetic field. Tests indicate positioning accuracy of the transport system to be within 0.13 mm.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Animals
  • Animals, Laboratory
  • Equipment Design
  • Magnetic Resonance Imaging / instrumentation*
  • Magnetic Resonance Imaging / veterinary
  • Models, Structural