[Assessment of left ventricular function with 99mTc-MIBI gated myocardial SPECT using 3 head rotating gamma camera]

Kaku Igaku. 1994 Jan;31(1):43-52.
[Article in Japanese]

Abstract

Twenty-one patients with ischemic heart disease were studied by resting 99mTc-methoxyisobutyl isonitrile (MIBI) gated SPECT images using a 3 head rotating gamma camera (Picker/Shimadzu: PRISM 3000). The data were acquired over 360 degrees in 20 (x 3) steps, each of which was 70 beats. The systolic wall thickening index (%WT) was calculated as (ES count-ED count/ED count x 100), using the circumferential profile analysis of ED and ES short axial images. The %WT was correlated well with LVEF obtained from contrast ventriculography (r = 0.85, n = 17). By using high speed (3 head) rotating gamma camera, we have shortened the time of collecting MIBI gated SPECT data, and this method can provide useful information on ventricular function.

Publication types

  • Case Reports

MeSH terms

  • Female
  • Gamma Cameras*
  • Gated Blood-Pool Imaging*
  • Humans
  • Male
  • Middle Aged
  • Myocardial Ischemia / diagnostic imaging
  • Myocardial Ischemia / physiopathology
  • Technetium Tc 99m Sestamibi*
  • Tomography, Emission-Computed, Single-Photon / instrumentation*
  • Ventricular Function, Left*

Substances

  • Technetium Tc 99m Sestamibi