Routine clinical monitoring of sirolimus (rapamycin) whole-blood concentrations by HPLC with ultraviolet detection

Clin Chem. 1996 Dec;42(12):1943-8.

Abstract

During phase I/II clinical trials of sirolimus (rapamycin; SRL), therapeutic drug monitoring was performed with a multistep liquid-liquid extraction of 1-mL aliquots of whole blood followed by reversed-phase HPLC with ultraviolet detection. Blood was sampled according to a standardized protocol and clinical status. SRL concentrations were interpolated from calibration curves with a linear range of 0-50 micrograms/L and 1 microgram/L lower limit of quantification. Quality control was monitored over 68 consecutive analytical runs by using frozen aliquots of SRL-supplemented pooled whole blood at 4, 12, and 32 micrograms/L. These samples showed mean concentrations of 3.7 +/- 0.6, 10.9 +/- 1.1, and 29.6 +/- 2.6 micrograms/L, respectively. This method for therapeutic drug monitoring of SRL permits one full-time technician to analyze 100 clinical specimens per week with a 24-h turnaround time. With this method, a strong linear relation (r2 = 0.946, Sy/x = 0.41, n = 115) between the average SRL concentration over a 24-h period and the SRL concentration at the 24th h was revealed.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Chromatography, High Pressure Liquid / methods*
  • Chromatography, High Pressure Liquid / statistics & numerical data
  • Drug Monitoring
  • Humans
  • Immunosuppressive Agents / blood*
  • Polyenes / blood*
  • Polyenes / pharmacokinetics
  • Quality Control
  • Sensitivity and Specificity
  • Sirolimus

Substances

  • Immunosuppressive Agents
  • Polyenes
  • Sirolimus