Comparison of usefulness of computer assisted continuous 48-h 3-lead with 12-lead ECG ischaemia monitoring for detection and quantitation of ischaemia in patients with unstable angina

Eur Heart J. 1997 Jun;18(6):931-40. doi: 10.1093/oxfordjournals.eurheartj.a015381.

Abstract

Aims: The selection of ECG leads used for ST monitoring may influence detection and quantitation of ischaemia.

Methods: We compared on-line continuous 48-h 12-lead against 3-lead ST monitoring in 130 unstable angina patients (Mortara. ELI-100). Onset and offset of ST episodes were defined by the lead with the first > or = 100 microV ST change relative to baseline and the lead with the latest return to baseline ST level, respectively. ST episodes were calculated for 12 leads and 3 leads (V2, V5, III) separately.

Results: ST episodes were detected in 88 patients (77%) by 12-lead and in 71 patients (62%) by 3-lead ST monitoring (P < 0.02). The median number (25.75%) of episodes/patient was 1 (0.3) for 3-lead and 2 (1.6) for 12-lead (P < 0.0001). The total duration of ischaemia detected during 12-lead far exceeded 3-lead monitoring: 12.3 (1, 58.2) and 1.7 (0, 23.3) min respectively (P < 0.0001). The probability of recurrent ischaemia declined most during the first 24 h of monitoring. After a period without ST changes of 1, 12, 24 and 36 h, the probabilities of recurrent ischaemia were 63, 31, 14 and 9%, respectively.

Conclusions: Continuous 12-lead ST monitoring increases detection rate and duration of ST episodes compared to 3-lead ST monitoring. The use of continuous 12-lead ECG monitoring devices on emergency wards and coronary care units is recommended.

Publication types

  • Clinical Trial
  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Angina, Unstable / complications*
  • Chi-Square Distribution
  • Diagnosis, Computer-Assisted*
  • Electrocardiography / instrumentation*
  • Electrocardiography / methods
  • Female
  • Humans
  • Male
  • Monitoring, Physiologic / methods
  • Myocardial Ischemia / diagnosis*
  • Myocardial Ischemia / etiology
  • Myocardial Ischemia / physiopathology
  • Sensitivity and Specificity