Feasibility of using finger arterial pressure in neonates

Arch Dis Child Fetal Neonatal Ed. 1997 Sep;77(2):F139-40. doi: 10.1136/fn.77.2.f139.

Abstract

The feasibility of using a Finapres device to reproduce the beat to beat signal of arterial blood pressure in eight neonates was assessed and compared with intra-arterial measurement of arterial blood pressure in the umbilical artery, using a catheter. The two methods gave similar results. Continuous recording of arterial blood pressure in neonates using Finapres is feasible and reliable.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Pressure / physiology
  • Blood Pressure Monitors*
  • Catheters, Indwelling
  • Evaluation Studies as Topic
  • Female
  • Humans
  • Infant, Newborn
  • Infant, Premature / physiology*
  • Intensive Care, Neonatal*
  • Male
  • Umbilical Arteries