Objectives: We sought to utilize terminal stored intracardiac electrograms (EGMs) to study the electrophysiologic events that accompany mortality in patients with third-generation implantable cardioverter-defibrillators (ICDs).
Background: Despite their ability to effectively terminate ventricular tachyarrhythmias, cardiac mortality in patients with ICDs remains high. The mechanisms and modes of death in these patients are not well understood.
Methods: We retrospectively analyzed clinical data and stored EGMs from patients enrolled in the clinical trial of the Ventritex Cadence ICD. Of the 1,729 patients 119 died during 6 years of follow-up. The final recorded EGM was reviewed. Postimplant EGMs as well as 50 control EGMs were used to define normal EGM characteristics.
Results: There were 36 noncardiac deaths (30%) and 83 cardiac deaths (70%). Of the cardiac deaths, 55 (66%) were nonsudden and 28 (34%) were sudden. When cardiac deaths were analyzed, 46 (55%) had no stored EGMs within 1 h of death, implying that the deaths were not directly related to tachyarrhythmias. In 37 cardiac deaths (18 nonsudden, 19 sudden), stored EGMs were present within 1 h of death. In these 37 deaths, the final EGM recorded was wide (>158 ms) in 33 (89%). Wide EGMs were interpreted as ventricular tachycardia in 27 and ventricular fibrillation in 6. In 13 of the 33 patients (39%) with wide EGMs, therapy was not delivered by the ICD, as it incorrectly detected a spontaneous termination of the arrhythmia. EGMs were significantly wider if recorded within 1 h, as compared with those recorded from 1 to 48 h before death (261+/-124 vs. 181+/-93 ms, p=0.04).
Conclusions: Only 37 patients (31%) who died after placement of an ICD had a stored EGM within 1 h of the time of death, suggesting that the majority of deaths (69%) were not the immediate result of a tachyarrhythmia. When EGMs were recorded, they were wide in 89% of patients. These wide EGMs most likely represent intracardiac recordings of electromechanical dissociation. Thus, of the 119 deaths, 112 (94%) were not the immediate result of a tachyarrhythmia.