Precision and accuracy of cross-sectional area measurements used to measure coronary endothelial function with spiral MRI

Magn Reson Med. 2019 Jan;81(1):291-302. doi: 10.1002/mrm.27384. Epub 2018 Jul 19.

Abstract

Purpose: Coronary endothelial function (CEF) reflects vascular health and conventional invasive CEF measures predict cardiovascular events. MRI can now noninvasively measure CEF by quantifying coronary artery cross-sectional area changes in response to isometric handgrip exercise, an endothelial-dependent stressor. Area changes (10 to 20% in healthy; 2 to -12% in impaired vessels) are only a few imaging voxels because of MRI's limited spatial resolution. Here, with numerical simulations and phantom studies, we test whether Fourier interpolation enables sub-pixel area measurement precision and determine the smallest detectable area change using spiral MRI.

Methods: In vivo coronary SNR with the currently used CEF protocol at 3T was measured in 7 subjects for subsequent in vitro work. Area measurements of circular vessels were simulated by varying partial volume, vessel diameter, voxel size, SNR, and Fourier interpolation factor. A phantom with precision-drilled holes (diameters 3-3.42 mm) was imaged 10 times with the current CEF protocol (voxel size, Δx = 0.89 mm) and a high-resolution protocol (Δx = 0.6 mm) to determine precision, accuracy, and the smallest detectable area changes.

Results: In vivo coronary SNR ranged from 30-76. Eight-fold Fourier interpolation improved area measurement precision by a factor 6.5 and 4.9 in the simulations and phantom scans, respectively. The current CEF protocol can detect mean area changes of 4-5% for SNR above 30, and 3-3.5% for SNR above 40 with a higher-resolution protocol.

Conclusion: Current CEF spiral MRI with in vivo SNR allows detection of a 4-5% area change and Fourier interpolation improves precision several-fold to sub-voxel dimensions.

Keywords: Fourier interpolation; MRI; coronary endothelial function; cross-sectional area measurement, vasodilation; spiral.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Atherosclerosis / diagnostic imaging
  • Computer Simulation
  • Coronary Circulation
  • Coronary Vessels / diagnostic imaging*
  • Endothelium, Vascular / diagnostic imaging*
  • Female
  • Fourier Analysis
  • Hand Strength
  • Humans
  • Magnetic Resonance Imaging*
  • Male
  • Middle Aged
  • Models, Theoretical
  • Phantoms, Imaging
  • Reproducibility of Results
  • Signal-To-Noise Ratio
  • Vasodilation